Owner manual
7 Maintenance and repairs
75 NIRFlex N-500 Operation Manual, Version F
Resetting the laser lifetime counter
After LASER unit replacement the following message appears:
Message concerning required instrument setup
Click the OK icon • to confirm the message.
Now click the Service icon • (Operator).
The following dialog opens:
Assembly setup dialog
Click the icon next to the Laser Module row • . The following dialog appears:
Exchange sequence dialog for the laser
Activate the watch icon next to the line • Laser hours of operation .
The following message box is displayed:
Message
Confirm the message by clicking the OK icon • .