3 Technical data
14 NIRFlex N-500 Operation Manual, Version F
3.2.2 Computer requirements
The spectral data are collected in the NIRWare database on the PC. The following minimum configura-
tion is needed for the PC running the NIRWare Software Suite (1.4):
Computer hardware and software requirements
Windows XP Professional (32-bit only), SP3 or higher
/ Windows 7 Professional / Ultimate (32-bit only)
Dual Core 2.4 GHz or faster
3 GB RAM or more (2 GB minimum)
10 GB of free harddisk space
CD– ROM drive
2 Network adapter (1 adapter minimum)
Display res.: 1280
x
1024 (1024
x
768 minimum)
Please consider a more powerful computer system if you are working with a bigger amount of data,
doing automated (cyclic) measurements or calculating your own calibrations.
3.3 Non-fiber optic measuring cells
3.3.1 NIRFlex Solids
Technical data NIRFlex Solids
Detector extended range InGaAs (temperature controlled)
Electric power supply 100 - 230 VAC ± 10%, 50/60 Hz, 20 W
Ambient conditions < 80 % relative humidity for T < 31 °C, linear decreasing to 67 % at
35 °C, max. 2500 m, for indoor use only
Ambient temperature 5−35 °C ( 25 ± 5 °C recommended)
Pollution degree 2
Installation category II
3.3.2 NIRFlex Solids Transmittance
Technical data NIRFlex Solids Transmittance
Detector InGaAs (temperature controlled)
Electric power supply 100 - 230 VAC ± 10%, 50/60 Hz, 20 W
Ambient conditions < 80 % relative humidity for T < 31 °C, linear decreasing
to 67 % at 35 °C, max. 2500 m, for indoor use only
Ambient temperature 5−35 °C ( 25 ± 5 °C recommended)
Pollution degree 2
Installation category II
Spectral range 12’500 - 6’000 cm
-1
(recommended range 11’520 - 6000 cm
-1
)
800−1660 nm (recommended range 870 - 1660 nm)
Photometric dynamic range 0 - 6 AU