Instruction Manual

NIRCal 5.5 Software Manual
152 NIRCal 5.5 Manual, Version A
The spectrum 2B is not a real outlier, but spectrum 1A is dangerous: it is better to select into the C-
Set.
3.17.2 Calibration Protocol Quantitative
Menu:
Calibration / Calibration Protocol Calculate or Show
Icon:
Short Key: F8
The calibration protocol is an important validation report giving all information about the data selection
and result of the chemometrics calculation. The calibration protocols are stored in the DB (or .nir
project file).
Explanation of the content of the calibration protocols:
Quantitative Calibration
The first section has the information about the user specific data selection and similar, as for the
Cluster calibration (row 1-58).