User Manual

5 Putting into operation
56 B-90 Operation Manual, Version B
!
Danger
Death or serious poisoning by gases or particles at O
2
sensor or filter malfunction
Exchange defective O
2
– sensor immediately
Exchange O
2
sensor regularly within the specified maintenance intervals
Exchange clogged filters immediately
Exchange filters regularly within the specified maintenance intervals
Safely dispose filter immediately
!
Warning
Death or serious injuries by compressed gas.
Depressurize gas circuit before any modification work
Wear safety goggles
!
Warning
Death or serious poisoning by contact or incorporation of condensed liquid.
Dispose of condensate safely after use
Wear safety gloves
Notice
Risk of instrument damage by wrong gas supply.
All labels at gas connections must be clearly ledgible
Before operation, check all gas connections for correct installation
External supply pressure must meet the system specifications
Notice
Risk of instrument damage by internal overpressure.
External supply pressure must meet the system specifications
Exchange clogged filters immediately
Safely dispose filter immediately