User Manual

2 Safety
12 B-90 Operation Manual, Version B
!
Danger
Death or serious poisoning by inhalation or incorporation of dried particles at recovery.
Wear safety goggles
Wear safety gloves
Wear a suitable protective mask
Wear a laboratory coat
Do not inhale dried particles
Stop drying gas flow before opening the drying circuit
Only recover particles in sufficiently ventilated flue or glove-box
Do not disperse the dried particles
Do not clean dusty parts with compressed air
!
Warning
Death or serious injuries by use in explosive environments.
Do not operate the instrument in explosive environments
Do not operate the instrument with explosive gas mixtures
Before operation, check all gas connections for correct installation
Directly withdraw released gases and gaseous substances by sufficient ventilation
!
Warning
Death or serious poisoning by contact or incorporation of harmful substances at use.
Before operation, check the instrument for correct assembling
Before operation, inspect sealings and tubes for good condition
Exchange worn out or defective parts immediately
Exchange clogged filters immediately
Only operate the instrument in ventilated environments
Directly withdraw released gases and gaseous substances by sufficient ventilation
Perform a dry-run without sample material and check for gas leakages