Instruction Manual
6 Operation
31 Melting Point M-565 Operation Manual, Version D
• Turn the rotary knob to
select a method.
• Press START to start the
determination.
• Use Edit to adjust method
parameters.
• Press the Options key to
get further functions:
• New: Creates a new
method.
• Delete: Deletes the
method.
• Rename: Changes the
method name.
6.2.5 Adjusting parameters during a determination
NOTE
Measuring parameters can be modified during preheating or a determination. This function is
possible only if enabled by the setting “Param. adj. during meas.” (see 6.5).
a
• Turn the rotary knob to
display the current param-
eter set.
• Adjust the parameters by
pressing Edit. Adjusting the
parameters does not affect
the saved method.
If the temperature gradient
is changed during a mea-
surement, the result shows
CHANGED (a) for the tem-
perature gradient.
6.2.6 Printout
NOTE
The result is printed out automatically if a printer is connected and automatic printing is configured in
the settings menu.