Instruction Manual

3 Technical data
14 Melting Point M-565 Operation Manual, Version D
3.2 Technical data overview
Table 3-3: Technical data of the Melting Point M-565
Melting Point M-565
Manual melting point determination
Manual boiling point determination
Automatic melting point determination
Automatic boiling point determination
Homogeneous sample loading
Positions for melting capillaries 3
Positions for boiling capillaries 1
Precision magnifying lens
Magnification of lens 2.5×
Digital camera
Video function
Magnification, display 6 x
Display Colour, TFT, 320×240, 3.5˝
Determination temperature range Ambient + 10 °C to 400 °C
Temperature resolution 0.1 °C
Accuracy melting point at 0.5 °C/min
up to 250 °C ± 0,3 °C
250 °C to 400 °C ± 0,3 °C to ± 0,5 °C
Repeatability melting point at 0.5 °C/min ± 0.1 °C
Accuracy boiling point at 1.0 °C/min up to
400 °C
± 0,5 °C
Repeatability b oiling point at 1.0 °C/min ± 0.3 °C
Temperature gradients, °C/min 0.1, 0.2, 0.5, 1, 2, 3, 5, 10, 20
Heat-up time (50 °C–350 °C) at 25 °C ~ 4 min
Cool-down time (350 °C–50 °C) at 25 °C ~ 13 min
Electrical supply 100–240 V (±10%), 50–60 Hz
Power consumption 150 W
Contact termination L, N, PE
Video run time 350 min at 1 °C/min, 700 min at 0.5 °C/min
Approval CE, CSA, UL
Dimensions (W×H×D), mm 190×200×370
Weight, kg 4.5
Environmental conditions For indoor use only
Temperature 5–40 °C
Altitude up to 2000 m a.s.l.
Humidity Maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C
Over voltage category II
Degree of protection IP 20
Pollution degree 2