Table 8-3: Error messages
Error number Possible cause Remedy
Error 01 Memory data loss, all data are reset. Calibrate device.
In case of recurrence contact customer service.
Error 02 Automatic restart, firmware problem possible. In case of recurrence contact customer service.
Error 03 Board temperature sensor defective. Contact customer service.
Error 04 Temperature sensor defective. Contact customer service.
Error 05 Heating defective. Contact customer service.
Error 06 Cooling fan defective. Contact customer service.
Error 07 Housing fan defective. Contact customer service.
Error 08 Camera defective, communication loss. Contact customer service.
Error 09 24 V fuse defective. Contact customer service.
Error 10 24 V input voltage missing. Contact customer service.
Error 11 Internal clock, power loss. Contact customer service.
Error 13 5 V input voltage missing Contact customer service.