6 Operation
35 Melting Point M-560 Operation Manual, Version D
6.4.4 Verification
To verify the calibration, proceed as follows:
• Measure all standards (BUCHI recommends using the verification set (11055019) but it is also
possible to use your internal standards.
NOTE
The verification is not a guided process.
• When all measuring results were within the required tolerances, click on Verified. The following
screen appears:
• Enter the following password: VER. The current date is now indicated under “Last verification”.
6.5 Settings, SysInfo, Test
Test: Functional test of the instrument. To
perform the test of several functions follow
carefully the on-screen instuctions. If a function
test does not show ok, please contact the cus-
tomer service. The Test protocol is described in
section 6.5.1.
SysInfo: Instrument gives information regarding
settings and connected devices. The SysInfo
protocol is described in section 6.5.2.
Settings: This menu can be used to change
parameters.