User Manual

6 Operation
33 Melting Point M-560 Operation Manual, Version D
• The instrument auto-
matically heats to the start
temperature of the first
substance. In the meantime,
prepare at least 6 samples of
the given substance. For this
process, follow section 6.2.1 .
Choose the correct phar-
macopoeia melting point
value with the rotary knob.
The pharmacopoeia melting
point is written on the
certificate of analysis. Press
Save.
To register temperature
stamps manually press the
Set button for the related
sample. Each button can
be pressed once. Regis-
tered temperature stamps
are displayed green.
After each run the current
result is indicated.
Press START to perform
next measurement.
After 6 results are obtained
within a standard deviation of
+/- 0.2 °C, the process moves
on to the next substance. The
process is identical for the
subsequent substances.
As soon as a complete
calibration is obtained, the
new calibration date is saved
automatically.
Press OK. All results
obtained are displayed.
The rotary knob can be
used to display all the
results of the calibration.
The Print key is used to
print out the data.
Back exits this menu.
The calibration date is now
saved and indicated on the
idle screen.