User Manual

6 Operation
29 Melting Point M-560 Operation Manual, Version D
Use the rotary knob to
choose the start tempera-
ture and press Next.
Use the rotary knob to
choose the stop tempera-
ture and press Next.
Use the rotary knob to
chose the temperature
gradient and press Save.
NOTE
Insert sample only when the instrument has reached the start temperature.
Press START to begin the
determination and follow
the instructions on the
screen.
• The instrument requests a
sample ID during preheating
(for details on entering text
see 6.1.3). This request can
be switched off by changing
the setting “Sample ID
request” (see 6.5).
Bar P: Enter the current
barometric pressure.
This request can be
switched off by changing
the setting “Barometric
pressure request” (see 6.5).
Insert sample and press
START to begin the deter-
mination.