User Manual
6 Operation
27 Melting Point M-560 Operation Manual, Version D
6.2.5 Adjusting parameters during a determination
NOTE
Measuring parameters can be modified during preheating or a determination. This function is
possible only if enabled by the setting “Param. adj. during meas.” (see 6.5).
a
• Turn the rotary knob to
display the current param-
eter set.
• Adjust the parameters by
pressing Edit. Adjusting the
parameters does not affect
the saved method.
• If the temperature gradient
is changed during a measure-
ment, the result shows
CHANGED (a) for the
temperature gradient.
6.2.6 Printout
NOTE
The result is printed out if a printer is connected.
a
b
c
d
e
Printout
a Instrument type
b Determination parameters
c Manual pressed results
d General information
e Signature of person making the
determination