7 Maintenance and repairs
61 B-290 Operation Manual, Version I
7.8 Replacing the oxygen sensor of the Inert Loop B-295
After approximately 2 years of operation the oxygen concentration displayed by the sensor decreases
drastically.
The sensor should now be replaced. For this purpose:
Open the front plate with an allen key.
Disconnect the cable by turning the plug.
Screw the sensor out and replace it by a new one.
Reconnect the cable.
Calibrate the sensor according to chapter 7.6, Cali-
brating the oxygen sensor.
Fig. 7.3: Replacing the oxygen sensor
7.9 Spray chilling accessory
Prior to cleaning, let the spray chilling accessory cool down with its covers on.
For cleaning, proceed as follows:
Fill the spray chilling accessory with the cleaning agent and open the needle valve.
If the accessory is blocked remove it.
Use hot water, soap solutions and a brush for cleaning.
Put the blocked metal parts in a heating oven to melt the stocked product.
Wash all parts with water and let them dry.
The silicone tubing between the spray chilling accessory and the nozzle has to be replaced if it is
blocked or broken.
7.10 Customer service
Only authorised service personnel are allowed to perform repair work on the instrument. These
persons have a comprehensive technical training and knowledge of possible dangers which might
arise from the instrument.
Addresses of official BUCHI customer service offices are given on the BUCHI website under:
www.buchi.com. If malfunctions occur on your instrument or you have technical questions or applica-
tion problems, contact one of these offices.
The customer service offers the following:
Spare part delivery
Repairs
Technical advice