If the system is run with a PTFE filter, the fine particles can par tially be recovered by counter-pulsating
the filter membrane and blowing the particles away. For this purpose, proceed as follows:
Fig. 6.5: Recovering particles of the outlet filter
Attach the PTFE membrane instead of the polyester filter and close the bottom with a polypro-
pylene plug.
Separate the manometer tube from the filter and remove the filter housing completely from the
holder.
Put the filter on the product feed table. Disconnect the nozzle cleaning tube from the nozzle and
connect it to the filter housing.
Close the outlet gas connection with an SVL 42 screw cap.
With the nozzle cleaning set on level 1, a pressure blow is generated every 5 seconds on the inner
side of the filter leading to a separation of the fines from the membrane. The particles fall into the
glass and can be recovered.