Instruction Manual

5 Putting into operation
39 B-290 Operation Manual, Version I
5.10 Quick hose couplings
Use
For fast and reliable connecting/disconnecting of polypress tubes e.g. the tubing of Inert Loop B-295
and Dehumidifier B-296.
Installation
Cut the polypress tube with a sharp knife or a special tube cutter into two halfs with a clean cut.
The tube ends must be straight!
Slip one hose clamp over each tube end.
Use one male and one female coupling adapter and install them at the tube ends.
Secure the coupling adapters with the hose clamps.
Test all tubes and connectors for tightness prior use!
Female coupling (including a flat seal)
Male coupling
Fig. 5.17: Quick hose couplings
5.11 Installing the Inert Loop B-295
!
Danger
Death or serious poisoning by gases or particles due to O
2
sensor or filter malfunction
Exchange defective O
2
– sensor immediately
Exchange O
2
sensor regularly within the specified maintenance intervals
Exchange clogged filters immediately
Exchange filters regularly within the specified maintenance intervals
Dispose of clogged or defective filters immediately