Instruction Manual

5 Putting into operation
37 B-290 Operation Manual, Version I
5.8 Tubing installation
The Mini Spray Dryer B-290 needs compressed gas for the two-fluid nozzle and the nozzle cleaner
between 5 and 8 bar. Therefore, the air or nitrogen is connected on the rear of the instrument via a
quick coupling.
Fig. 5.13: Quick coupling rear side
The Mini Spray Dryer B-290 can be used in an open or closed mode. The open mode is set in the
suction mode as a standard. However, if the exhaust gas is aggressive and could lead to a cor ro sion
of the aspirator, the instrument can also be run in a pressure mode. This mode leads to a higher noise
emission.
NOTE
Risk of instrument malfunction when the instrument is operated with bent hoses.
Always inspect the instrument for bends or kinks in hoses and eliminate them prior to operation.
Exchange hoses with signs of mechanical stress such as small cracks and blind spots at the
surface immediately.
1
2
a Tubing between cyclone respectively filter and
aspirator
b Exhaust tube
Fig. 5.14: Open mode set-up in suction mode