5 Putting into operation
29 B-290 Operation Manual, Version I
5.4 Installing the glass assembly
!
CAUTION
Risk of minor or moderate cuts by sharp edges.
Do not touch defective or broken glassware with bare hands
Do not touch thin metal edges
To install the spray cylinder, proceed as follows:
1
2
3
4
7
6
8
5
Screw the seal holder b on the instrument
using the three knurled screws.
Connect the separation flask d to the spray
cylinder c by closing the screw connection
(SVL 42) tightly.
Screw the connection piece f on the cylinder
outlet flange with the left flange screw joint g.
The connection is sealed with a Viton o-ring.
To mount the cylinder on the instrument,
hold the glass in the upper flange around the
nozzle and insert the connection piece into
the support element h.
Open the fixation e.
Press the glass flange on the above seal.
Push the lever a carefully on the left side, so
that the flange of the cylinder is held firmly.
Close the fixation e.
12
11
13
4
9
6
14
10
Attach the cyclone k together with the
product collection vessel n to the connection
piece f with a second flange screw joint j.
Screw the glass elbow i onto the top of the
cyclone.
Connect the grounding cable l to the cover
of the product collection vessel m and to the
housing to minimize the electrostatic charge
of the powder.