User Manual
7 Maintenance and repairs
45 B-811 Operation Manual, Version G
7.4 Magnetic glass valve
5
1
3
4
2
6
a Cover cap
b Glass valve
c Valve housing
d Anchor
e Spring
f Inductor coil
Fig. 7.1: Exploded view of the glass valve
Insert the spring • e into the anchor d.
Screw the valve housing • c and the magnetic valve together f.
Push the valve onto the glass part • b.
Screw the cover • a on and tighten slightly.
7.5 Membrane assembly
Insert anchor with membrane at an angle into •
valve housing
Guide entirely through opening by pressing •
slightly with circular movements.
Fig. 7.2: Membrane assembly
7.6 Heating elements
The heating elements should be cleaned regularly after they have cooled down. Remove dirt adhering
to the plate with e.g. ethanol. Then clean the plate with a moist cloth and dry it.
ATTENTION
Never use scratching sponges, cleaning pads or cleaner containing abrasives to clean the heating
elements as they might damage the surface.