User's Manual

243
To create a new strength protocol, click New Test and then click Integrated Isometric Strength Protocol.
Enter in a protocol name and select the FOCUS for the Integration type. Generally for a strength test the
trial duration is 3 to 5 seconds in length with the rest between trails set to 5 seconds to allow the client’s
energy stores to build up again. A rest period of 5 seconds usually gives the evaluator enough time to set
the client up for testing on the other side. The Custom Fields allow for additional information to be
It should be noted that once the Start button is clicked, a message will appear
stating that the force meter needs to be zeroed. This is seen in the diagram
below. As there are many attachments for each load cell, ODES¥ will account
for the weight of the accessory attachment during each test. No force should
be applied to the force meter.