User's Manual
158
Elbow Flexion
With the client in a standing position, manually landmark the lateral forearm with the hand in a supinated
position. Using only the Master sensor, align it in the sagittal plane. Click the button on the Master sensor
to take the initial reading. Have the client maximally flex the elbow. Click the button on the Master sensor
a second time to take the final reading. ODES will automatically document the final true range. Return the
client's elbow to the neutral position. Repeat these steps two more times on the same side and then
repeat 3 times for the other side.
Wrist Pronation
With the client in a standing position, manually landmark the distal radial head with the wrist in a neutral
position. Using only the Master sensor, align it in the coronal plane. Click the button on the Master sensor
to take the initial reading. Have the client maximally pronate the wrist. Click the button on the Master
sensor a second time to take the final reading. ODES will automatically document the final true range.
Return the client's wrist to the neutral position. Repeat these steps two more times on the same side and
then repeat 3 times for the other side.