User's Manual

294
You are now able to modify the number located beside ‘F’ to ‘6’. To do this, click on the number 3 until it
reads 6. You have now set the Algometer to read in only port 6 for both strength and fibromyalgia tests.
Ensure that the Algometer is plugged into port 6.
Troubleshooting the Multiplanar Accessory Housing Joint
If you have problems inserting or removing the Multiplanar Accessory Housing Joint from one of the ER
arms, try the following:
Locate the small round silver pin on the right side of the upper or lower accessory arm (this pin is located
at the end of each arm). Try pulling the pin outward (further to the right), and then twist it slightly in any
direction. This will create space in the opening of the accessory arms’ end.
Locate the short black handle on the right side of the upper or lower accessory arm. This handle is
located at the end of each accessory arm. Turn the handle in a counterclockwise direction. This will
loosen the opening in the accessory arm.
These steps should loosen the openings in the ends of the accessory arms, and create space for the
insertion or removal of the Multiplanar Accessory Housing Joint.
If you have problems securing the Multiplanar Accessory Housing Joint to one of the ER arms, try the
following:
Locate the small round silver pin on the right side of the upper or lower accessory arm. This pin is
located at the end of each arm. Ensure that the pin is in the locked position (tucked into the far left). If it
is not, twist it slightly in any direction, until you hear a click. This will tighten the space in the opening of
the accessory arm’s end and secure the Multiplanar Accessory Housing Joint.
Troubleshooting: Accessory Attachment and Removal
If an accessory cannot be inserted into the Accessory Housing Joint, try the following:
Locate the handle on the top of the Accessory Housing Joint. This handle is on the right side of the
Multiplanar Accessory Housing Joint. Make sure that the handle is set to the unlocked position. In order
to be in the unlocked position, the handle must be given a ¼ turn in a counter-clockwise direction, and
then pulled outward. The handle can be maintained in the unlocked position by twisting it gently in either
direction.
Select the accessory to be attached. Be sure that the flattened end of the accessory is being inserted
into the housing mechanism facing to the right.
Grip the handle and twist it gently until a click is heard.
If an accessory cannot be removed from the accessory housing mechanism, try the following:
Locate the top handle of the Multiplanar Accessory Housing Joint.
Pull the handle and twist it gently in any direction.