User's Manual
291
Hand Grip, Pinch Grip, Algometer, and FOCUS
“No Testing can be done until this device has been calibrated or verified”:
Has the tools been calibrated and verified? If not, calibrate and verify.
Is the Calibration/Verification reminder set to “must be done” or “reminder” in Administration |
Environment Settings? If on “must be done”, calibration must occur prior to performing a test. If set to
“reminder” tests can be performed without a new calibration.
The voice is not beginning the count down on your screen. You cannot perform any test (no prompts).
Go into Administration ŇEnvironment Settings. Make sure that Voice Response is on. Try to switch the
voice type to see if there are any problems with the speech software. If an error occurs you will need to
re-load the SAPI51 file located on the ODES CD under the Speech folder.
Cannot start the strength test that has been selected
Are your voices installed properly?