Recommended Care and Maintenance Schedule
The following outlines the recommended maintenance schedule for your Hanoun System. A maintenance
log should be kept for the equipment.
It is recommended that if your computer is going to be used for other applications than ODES or if the
computer is connected to a network or the Internet, anti-virus software be installed.
After each Client
x Clean handles with an antibacterial wipe or rubbing alcohol (70% Alcohol) for sanitary reasons.
Daily
x Verify the equipment to be used to ensure it is measuring accurately. If verification fails, re-
calibrate the equipment.
x Make sure Handgrip is stored on the computer management stand with the end with the cord in
the top hook.
x Store all equipment so that the weight of the tool is not on the cord.
x Turn off the data acquisition box and computer at the end of the day.
x Back up your database (have a backup disk for each day of the week)
Weekly
x Calibrate and verify all tools.
x Back up your database (have a back up disk for each week of the month)
Monthly
x Clean the ER Column with a mild glass (15-20% Alcohol) or degreasing cleaner with a lint free
cloth.
x Re-lube the column with lightweight general lubrication oil using a lint free cloth.
x Check all wires to ensure they are secure and in good condition.
x Compact and Repair the ODES database. Refer to ODES section of manual.
x Clear hard drive of unnecessary files by going to Start ŇPrograms ŇAccessories ŇSystem Tools Ň
Disk Cleanup. Select the C Drive and select the files you wish to delete.
x Run the defragmenter program to ensure optimum computer performance. To do this go to Start
ŇPrograms ŇAccessories ŇSystem Tools Ň Disk Defragmenter.
Quarterly
x Remove all handles from the ER arms, FOCUS load cell and UTM and apply a white grease to
them.
x Check all the bolts on the EPIC crate and lifting shelves to make sure they are secure
Yearly
x Make sure the ER Platform is still level and make adjustments as appropriate.