User's Manual
64
Enter the amount of weight that will be used to calibrate the left and right force scales in the appropriate
fields before you click OK. You must use a minimum of 25 pounds and for convenience use the same
weight for the right and left side. Calibrate one load cell, and then the other.
Once all of the tasks in the calibration have been completed, and no problems arose, the software will ask
you to enter the name of the individual who performed the calibration.
Weight Verification
After calibrating the tool through either automatic or manual calibration, it is recommended that the
accuracy of the device be verified. Verification should be completed prior to testing each day or after the
equipment has been set up. To verify go to Calibration | Verify Calibrations in the top task bar; or from
Verify in the bottom right-hand corner of the Calibration page, after calibration has been completed.
Just before calibration of the Force Measurements, shown in the screen below,
make sure to remove the left and right handles from the UTM.