User's Manual

52
Please ensure that the Hand Grip is plugged into PORT 4 on the Data Acquisition Box.
After selecting the manual calibration option and the tool you wish to calibrate, the calibration page will
display as follows.
The above screen shows the last day of successful calibration (in this case it was April 17th), the type of
calibration and the directions involved in manual calibration. In order to calibrate, first locate the correct
attachment for the Hand Grip.
The first step in manual calibration of the Hand Grip is to place the tool in the square base
attachment.Click OK in order to zero the Hand Grip.
Next, attach the ring-shaped calibration tool to the rungs on the Hand Grip dynamometer (leave the hand
grip handle attached) while the tool is lying on its back. Notice that the attachment connects to the top
rung of the Hand Grip on one side and the second rung from the top on the other. This creates a flat
surface for weight to be placed upon. The calibration tool weight s 0.3 pounds.