Owner's manual
2-4
Brooks
®
Model 5860i
Section 2 Installation
Installation and Operation Manual
X-TMF-5860i-MFM-eng
Part Number: 541B110AAG
November, 2008
Figure 2-1 Model 5860i Dimensions
Connection "X" Dim.
1/4" Compression
4.32
Fitting 109.7
1/4" Tube VCO
3.86
98.0
1/4" Tube VCR
4.18
106.2
3/8" Compression
4.44
Fitting 112.8
Figure 2-2 "D" Type Connector Pin Arrangement
PIN NO. FUNCTION COLOR CODE
1* Set Point Return Black
2 Volt Signal Output White
3 Not Used
4 Current Signal Output Green
5 +15 to +28 Vdc Supply Orange
6 Not Used
7* Current Setpoint Input Wht/Blk
8 Voltage Setpoint Input Red/Blk
9 Supply Common Grn/Blk
10 Signal Output Return Org/Blk
11* +5 Volt Reference Output Blu/Blk
12* Valve Override Input Blk/Wht
13 Not Used
14 Chassis Ground Grn/Wht
15 Not Used
* These connections are used for the 5850i and 5851i Mass Flow Controllers.
Make no connections to these pins.
NOTE:
1. Cable shield tied to chassis ground
in meter connector. Make no
connection on customer end.
2. All power leads must be connected
to power supply.