Owner's manual
2-3
Brooks
®
Model 5860i
Section 2 Installation
Installation and Operation Manual
X-TMF-5860i-MFM-eng
Part Number: 541B110AAG
November, 2008
2-4 Gas Connections
Standard inlet and outlet connections supplied on the Model 5860i are1/
4" compression fittings for flow rates up to 10 slpm, and 3/8" compression
fittings for higher flow rates.
Optional 1/8 compression, VCO
TM
and VCR
TM
connections are available
upon request. Prior to installation, make certain all piping is clean and free
of obstructions. Install the piping in such a manner that permits easy
removal if the instrument is to be removed for cleaning or test bench
troubleshooting.
2-5 Installation (Refer to Figures 2-1 through 2-2)
Recommended installation procedures:
a. The Model 5860i should be located in a clean dry atmosphere
relatively free from shock and vibration.
b. Leave sufficient room for access to the electrical components.
c. Install in such a manner that permits easy removal if the instrument
requires cleaning.
d. The Model 5860i Mass Flowmeter can be installed in any position.
However mounting orientations other than the original factory
calibration (see data sheet) will result in a ±0.5% maximum full scale
shift after re-zeroing.
e. When installing flowmeters with full scale flow rates of 10 slpm or
greater, be aware that sharp, abrupt angles in the system piping
directly upstream of the meter may cause a small shift in accuracy.
If possible have at least 10 pipe diameters of straight tubing up-
stream of the Model 5860i Mass Flowmeter.