Figure 2-1 Model 5860E Dimensions
Connection "X" Dim.
1/4" Compression
4.32
Fitting 109.7
1/4" Tube VCO
3.86
98.0
1/4" Tube VCR
4.18
106.2
3/8" Compression
4.44
Fitting 112.8
Figure 2-2 "D" Type Connector Pin Arrangement
PIN NO. FUNCTION COLOR CODE
1 Cmd. Common (Command Pot "CCW") • Black
2 0-5 Volt Signal Output White
3 Supply Common Red
4 Valve Off • Green
5 +15 Vdc Supply Orange
6 -15 Vdc Supply Blue
7 Valve Test Point/Purge • Wht/Blk
8 Cmd. Input or Cmd. Pot "S" • Red/Blk
9 Supply Voltage Common Grn/Blk
10 0-5 Volt Signal Common Org/Blk
11 +5 Volt Reference Output (Command Pot "CW") • Blu/Blk
12 Valve Override • Blk/Wht
13 Not Used Red/Wht
14 Chassis Ground Grn/Wht
15 Remote Transducer Input* • Blu/Wht
* Factory Activated Option.
• These connections are used for the 5850E and 5851E Mass Flow Controllers.
Make no connections to these pins.
NOTE:
1. Cable shield tied to chassis ground in
meter connector. Make no connection
on customer end.
2. All power leads must be connected to
power supply.