User's Manual
Table Of Contents
- 1 Identification
- 2 Declaration of Conformity
- 3 General Instructions
- 4 Safety Instructions
- 5 Product Specifications
- 6 Installation
- 7 Operation
- 7.1 Operating personnel
- 7.2 Protocol change
- 7.3 Operation of the ASC-I1 protocol
- 7.3.1 Structure of the communication protocol
- 7.3.2 Package contents
- 7.3.3 Commands of protocol
- Commands host to reader
- Commands reader to host
- Message items
- X - Read data
- W - Write data
- R - Automatic read
- G - Request parameter value
- P - Set parameter
- N - Reset
- e - Error message
- H - Heartbeat
- V - Query software version
- L - Lock a page of the transponder
- I - Set tuning of the RF module
- J - Request tuning settings of RF module
- A - Sensor event
- 7.3.4 Parameters
- 7.3.5 Message examples
- 7.4 Operation of SECS/HSMS Protocol
- 7.4.1 Introduction
- 7.4.2 SECS-I implementation
- 7.4.3 HSMS option
- 7.4.4 HSMS implementation
- 7.4.5 SECS-II implementation
- 7.4.6 Parameters
- Parameter 0 (0x00)
- Parameter 1 (0x01)
- Parameter 2 (0x02)
- Parameter 3 (0x03)
- Parameter 4 (0x04)
- Parameter 5 (0x05)
- Parameter 6 (0x06)
- Parameter 7 (0x07)
- Parameter 8 (0x08)
- Parameter 9 (0x09)
- Parameter 11 (0x0B)
- Parameter 12 (0x0C)
- Parameter 20 (0x14)
- Parameter 22 (0x16)
- Parameter 23 (0x17)
- Parameter 24 (0x18)
- Parameter 25 (0x19)
- Parameter 26 (0x1A)
- Parameter 27 (0x1B)
- Parameter 29 (0x1D)
- Parameter 30 (0x1E)
- Parameter 31 (0x1F)
- Parameter 32 (0x20)
- Parameter 33 (0x21)
- Parameter 34 (0x22)
- Parameter 35 (0x23)
- Parameter 36 (0x24)
- Parameter 37 (0x25)
- Parameter 38 (0x26)
- Parameter 40 (0x28)
- Parameter 41 (0x29)
- Parameter 42 (0x2A)
- Parameter 43 (0x2B)
- Parameter 44 (0x2C)
- Parameter 45 (0x2D)
- Parameter 98 (0x62)
- Parameter 99 (0x63)
- 7.4.7 SEMI E99
- 7.4.8 Message details
- 7.4.9 HSMS message examples
- 7.4.10 SECS-I message examples
- 8 Service and Troubleshooting
- 9 Dismantling and Storage
- 10 Transport and Disposal
- 11 Accessories
Product Manual - RFID Reader LF60C SoliD 152
Chapter 8
Service and Troubleshooting
8.4.3 Stream function
TE Tag error
no transponder or
antenna installed
no readable tag within
the reading range
put a transponder into the antenna
range, verify type and function of
the transponder
antenna is not
connected correctly
check antenna connection
antenna and
transponder are in a
bad orientation
check orientation between antenna
and transponder (see
Antenna
installation)
antenna is not tuned perform an antenna tuning
disturbing field at
transmitting freqency
check antenna surroundings for
possible disturbing sources
(monitors, servo motors, ...)
antenna is damaged or
too close to metal
exchange antenna, verify antenna
installation
SSACK Name Description Possible cause Correcting action
Stream
Function
Description Possible cause Correcting action
S9F1 unrecognized DeviceID
message with wrong Devi-
ceID was sent to the reader
send a message with the correct
DeviceID (can be taken from the
S9F1 message)
S9F3 unrecognized stream type
message with a unkown
stream type was sent to the
reader
check stream function syntax of the
message
S9F5 unrecognized function type
messagw with a unkown
function type was sent to the
reader
check strem function syntax of the
message
S9F7 illegal data wrong RIC at reset message check RIC value
S9F9 Transaction Timer Timeout
host system does not
acknowledge the message or
sends no answer
Check availability of the host sys-
tem (terminal), check RS232 cable
and connector, check Ethernet
cable and connector, check link LED
and traffic LED, check IP address
settings