User's Manual
Table Of Contents
- 1 Identification
- 2 Declaration of Conformity
- 3 General Instructions
- 4 Safety Instructions
- 5 Product Specifications
- 6 Installation
- 7 Operation
- 7.1 Operating personnel
- 7.2 Protocol change
- 7.3 Operation of the ASC-I1 protocol
- 7.3.1 Structure of the communication protocol
- 7.3.2 Package contents
- 7.3.3 Commands of protocol
- Commands host to reader
- Commands reader to host
- Message items
- X - Read data
- W - Write data
- R - Automatic read
- G - Request parameter value
- P - Set parameter
- N - Reset
- e - Error message
- H - Heartbeat
- V - Query software version
- L - Lock a page of the transponder
- I - Set tuning of the RF module
- J - Request tuning settings of RF module
- A - Sensor event
- 7.3.4 Parameters
- 7.3.5 Message examples
- 7.4 Operation of SECS/HSMS Protocol
- 7.4.1 Introduction
- 7.4.2 SECS-I implementation
- 7.4.3 HSMS option
- 7.4.4 HSMS implementation
- 7.4.5 SECS-II implementation
- 7.4.6 Parameters
- Parameter 0 (0x00)
- Parameter 1 (0x01)
- Parameter 2 (0x02)
- Parameter 3 (0x03)
- Parameter 4 (0x04)
- Parameter 5 (0x05)
- Parameter 6 (0x06)
- Parameter 7 (0x07)
- Parameter 8 (0x08)
- Parameter 9 (0x09)
- Parameter 11 (0x0B)
- Parameter 12 (0x0C)
- Parameter 20 (0x14)
- Parameter 22 (0x16)
- Parameter 23 (0x17)
- Parameter 24 (0x18)
- Parameter 25 (0x19)
- Parameter 26 (0x1A)
- Parameter 27 (0x1B)
- Parameter 29 (0x1D)
- Parameter 30 (0x1E)
- Parameter 31 (0x1F)
- Parameter 32 (0x20)
- Parameter 33 (0x21)
- Parameter 34 (0x22)
- Parameter 35 (0x23)
- Parameter 36 (0x24)
- Parameter 37 (0x25)
- Parameter 38 (0x26)
- Parameter 40 (0x28)
- Parameter 41 (0x29)
- Parameter 42 (0x2A)
- Parameter 43 (0x2B)
- Parameter 44 (0x2C)
- Parameter 45 (0x2D)
- Parameter 98 (0x62)
- Parameter 99 (0x63)
- 7.4.7 SEMI E99
- 7.4.8 Message details
- 7.4.9 HSMS message examples
- 7.4.10 SECS-I message examples
- 8 Service and Troubleshooting
- 9 Dismantling and Storage
- 10 Transport and Disposal
- 11 Accessories
Product Manual - RFID Reader LF60C SoliD 149
Chapter 8
Service and Troubleshooting
8.4 Error codes
8.4.1 Error ID and Alarm ID
ID Name Description Possible cause Correcting action
0 none no error not used no
1 auto fail automatic reading is not
possible *
reader is still busy with
a former read or write
request
wait until the former request is
done
2 ex fail read or write initiated
from the host and/or
other actions cannot be
carried out *
reader is still busy with
a former read or write
request
wait until the former request is
done
3 write fail data transfer to the
transponder not
possible *
reader is still busy with
a former read or write
request
wait until the former request is
done
4 no tag
no transponder or
antenna is installed
no readable tag within
the reading range
put a transponder into the antenna
range, verify type and function of
the transponder
antenna is not
connected correctly
check antenna connection
antenna and
transponder are in a
bad orientation
check orientation between antenna
and transponder (see
Antenna
installation)
antenna is not tuned perform an antenna tuning (see
DIP switches)
disturbing field at
transmitting freqency
check antenna surroundings for
possible disturbing sources
(monitors, servo motors, ...)
antenna is damaged or
too close to metal
exchange antenna, verify antenna
installation
5 invalid
invalid parameter or
data
data sent with a
command are wrong
check syntax and data of command
sent parameter is not
implemented or is out of
range
check syntax and value of
parameter
6 unknown unknown error not used no