Transponder Reader LF60 Gen2 (SECS1-Protocol), Release 0.2
Parameter 20: sensor delay for presence sensor
Delay time for sensor signal to start an automatic read.
0 .. 255 1/10 s
Default: (10) 1s
Parameter 21: Not defined!
Parameter 22: sensor-triggered action for presence sensor
0 : Read all transponders
1 : Read page 1 of a multipage transponder
...
17 : Read page 17 of a multipage transponder
240 : Read a read/only transponder
241 : Read a read/write transponder
Default : (0) read all transponders
Parameter 23: triggered read frequency
If a read/write error occurs, the triggered read frequency defines the
time between two attempts to read or write a transponder; or the read
frequency if there is a triggered read (no polling).
2 .. 10 from 1s
Default: (5) 500ms
Parameter 24: r/w maxrepeat
If a read/write error occurs, this parameter defines the maximum
number of attempts to read or write a transponder.
0 .. 255
Default: 5