User's Manual
Table Of Contents
- 1 Identification
- 2 Declaration of Conformity
- 3 General Instructions
- 4 Safety Instructions
- 5 Product Specifications
- 6 Installation
- 7 Operation
- 7.1 Operating Personnel
- 7.2 Theory of Operation
- 7.3 Data Items Dictionary
- 7.4 Attribute (ECID and SVID) Values
- AlarmStatus (Read only)
- ASCII_T1 (ASCII only)
- ASCII_T3 (ASCII only)
- BAUDRATE
- CarrierIDLength
- CarrierIDOffset
- CHECKSUM (ASCII only)
- CID_DISPLAY
- CID_E99_PAD
- CID_ERROR
- CID_JUSTIFY
- CID_MAX_LENGTH
- CID_NP_ASCII
- CID_PAD
- Configuration (Read only)
- DeviceID
- DeviceType (Read only)
- DUAL_SENSOR
- ENABLE_EVENTS
- ENABLE_TIMEOUTS (ASCII only)
- EXTENDEDSSACK
- HardwareRevision- Level
- HeadID
- HeadStatus
- HOST_CONT_PORT1 _LED
- HOSTNAME
- MANTWRITEONLY
- Manufacturer
- MDLN
- ModelNumber
- OperationalStatus
- PARITY
- PIP (Read only)
- PIP_AUTOREAD
- PIP_AUTOREAD_ DATA
- PIP_AUTOREAD_ LENGTH
- PIP_SENSOR_ POLARITY
- RADIO_RETRY
- RDA
- RW_ADJUSTMENT
- RW_REPEATTIME
- SELF_TEST_ RESULT
- SENSOR_TIMEOUT
- SERIALNUM
- SerialNumber
- SIGNALSTRENGTH (Read only)
- SW_PARTNUMBER
- SOFTREV
- SoftwareRevision- Level
- STATUS_ENABLE
- TARGETID
- USETESTDIP
- 7.5 Operation of the SECS Protocol
- 8 Service and Troubleshooting
- 9 Dismantling and Storage
- 10 Transport and Disposal
Brooks Automation
260301 Revision A 75
ATR60LF RFID Reader CAN Bus 7 Operation
Product Manual 7.5 Operation of the SECS Protocol
S18F11: WRITE MATERIAL ID REQUEST (WMID) (H -> E)
This message is used to request the subsystem indicated by the TARGETID
to write the MID.
S18F11 W
L,2
<TARGETID>
<MID>
S18F12: WRITE ID ACKNOWLEDGE (WIA) (E -> H)
This message acknowledges the success or error of writing the MID to the
subsystem indicated in the TARGETID.
S18F12
L,3
<TARGETID>
<SSACK>
L,s
<STATUS1>
…
<STATUSs>
If the TARGETID is unknown, a communication error (CE) occurs.
Exceptions:
For single-page TIRIS transponders, only 8 bytes are written.
If this command fails, some or all of the previous data in the
transponder might be lost. Read the transponder to confirm the
contents of the transponder.
If the TargetID is invalid, no statuses are sent (list of zero).
ATTENTION
The reader must be in the maintenance mode to write the MID with
message S18F11.
ATTENTION
If the length of the MID is longer than what the subsystem can accept,
the subsystem will return a failure message in the reply SSACK. If the
length of the MID is less than the maximum allowed length of device,
then the MID will be padded with “0” in the front. Refer to MicroTag
READ/WRITE OPERATION in this document.