Manual

Brookfield Engineering Labs., Inc. Page 19 Manual No. M/98-206-A0402
The actual response to the above command will be RS1004F2{CR}. The last digit indicates that
the system is in stand-by mode (see Table 8).
4) To change the operating mode, enter
RA{s
t
}{CR} at the external device where s
t
indicates the
desired state of the controller (as described in Table 8) and
must be either a 1 or a 2.
s
t
Codes Defining the State of the Programmable Temperature Controller
Code Definition
1 RUN mode; control to the setpoint temperature
2 STANDBY mode; do not control; wait for further input
3 Probe error 1 (probe disconnected from controller)
Overheat of the Thermo Container. This occurs after approximately 3
4 minutes of constant heat output with no sensed temperature rise in the
Thermo Container.
5 Probe error 2 (temperature reading is above allowable limit)
TABLE 8
NOTE: Only codes 1 and 2 will be accepted by the Programmable Temperature Controller.
Codes 3, 4 and 5 are indications only. All five codes may be returned by the Pro-
grammable Temperature Controller to the connected device (i.e. PC or terminal).
Any command other than those listed above sent to the Programmable Temperature Controller
will cause the controller to respond with a question mark (?{CR}).
NOTE: Temperature/Time programs entered in stand-alone mode (i.e. from the Programmable
Temperature Controller front panel) cannot be executed in the external mode via the RS-
232C port. The commands explained above are the only means with which to control the
Programmable Temperature Controller.
VI. REMOTE OPERATION USING RHEOCALC
©
RHEOCALC
©
, a software program available from Brookfield Engineering Laboratories that controls
the operation of the Brookfield DV-III+ Rheometer, can also be used concurrently to control the
Programmable Temperature Controller in its external mode.
VII. ERROR MESSAGES AND FAILSAFES
If the display shows an error message or there is a continuous “beeping” from the piezo buzzer, you
are in one of the following error conditions:
1) The probe has become disconnected from the Programmable Temperature Controller. Plug the
probe back in to the Programmable Temperature Controller to correct the error.