MODEL 30 pH/DO SECTION 5.0
PROGRAMMING THE ANALYZER
5.5.4 Procedure.
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
1. Press MENU. The main menu screen appears. Choose Program.
2. Choose Measurement.
3. Choose Sensor 1 (DO) or Sensor 2 (pH). For a single input configura-
tion, the Sensor 1 Sensor 2 screen does not appear.
If you chose Sensor 1, go to step 4.
If you chose Sensor 2, go to step 11.
4. For Sensor 1 (DO), identify the manufacturer of the sensor, Rosemount
Analytical or other. Choose other for Broadley-James sensors.
5. Identify the application. Biopharm is biopharmaceutical manufacturing.
Other is aeration basins, boiler feedwater, etc.
6. Select the desired units. The screen at left appears only if Rosemount
and Other were selected in steps 4 and 5. Otherwise, the choice ppb
does not appear.
7. Choose the desired pressure units: mmHg, inHg, atm, kPa, or bar.
8. Choose the amount of filtering desired.
9. Enter the salinity (in part per thousand of sodium chloride) in the sample.
10. The display returns to the screen shown in step 3. To configure the pH
(ORP) sensor, choose Sensor 2. To return to the previous screen, press
EXIT. To return to the main display, press MENU followed by EXIT.
11. If Sensor 2 (pH) was selected, the screen at left appears. Select pH,
Redox, or ORP. If pH was selected, go to step 12; otherwise, go to step 17.
12. Choose No if the pH sensing electrode is NOT a glass electrode.
13. Choose Soln Temp Corr or Sensor Isoptntl.
14. For Soln Temp Corr, choose Off, UltraPure, HighpH, or Custom.For
Custom, enter the desired temperature coefficient.
15. For Sensor Isoptntl, enter the desired sensor isopotential pH. Do not
change the sensor isopotential pH unless the sensor is known to have an
isopotential pH different from 7.00.
16. The display returns to the screen shown in step 3.
17. If Redox or ORP was selected, there are no further settings to make. The
display returns to the screen shown in step 3. Press EXIT to return to the
previous screen. To return to the main display, press MENU followed by
EXIT.
Calibrate Hold
PPrrooggrraamm
Display
Configure?
SSeennssoorr11
Sensor2
S1Manufacturer?
RRoosseemmoouunntt
Other
S1Application?
Biopharm
OOtthheerr
S1 units in?
ppppmm
ppb %sat
Pressure units?
mmmmHHgg
inHg >>
Input filter?
63% in
00
05sec
Salinity, parts/
thousand?
22
0.0%.
S2Measure?
ppHH
Redox ORP
S1 Glass Fault?
Enable? Yes No
S1 SolnTempCorr?
OOffff
Ultrapure >>
Sensor Isoptntl
S1:
00
7.00pH
Outputs Alarms
MMeeaassuurreemmeenntt
>>
SSoollnn TTeemmpp CCoorrrr
Sensor Isoptntl
25