Specifications

16
MODEL 30 pH/DO SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS
1. SENSOR-OUTPUT ASSIGNMENTS (type of dissolved oxygen measurement, pH, ORP, or redox is selected
during Quick Start)
3. OUTPUT RANGES (type of dissolved oxygen measurement, pH, ORP, or redox is selected during Quick Start)
Measurement Range Section
Dissolved oxygen - ppm 0 to 20 ppm 5.3
Dissolved oxygen - ppb 0 to 900 ppb 5.3
Dissolved oxygen - % sat. 0 to 200% 5.3
pH 0 to 14 5.3
ORP/Redox -1400 to 1400 mV 5.3
Temperature 0 to 100°C 5.3
Sensor(s) Output 1 Output 2 Section
Single sensor Dissolved oxygen Temperature 5.3 and 5.9
Dual sensor Dissolved oxygen (sensor 1) pH/ORP/Redox (sensor 2) 5.3 and 5.9
2. OTHER OUTPUT SETTINGS
Output Dampening 0 or 4 mA Mode Section
1 off 4 Linear 5.3
2 off 4 Linear 5.3
4. ALARM CONFIGURATION AND SETPOINTS
Alarm
1 2 3 Section
Assigned to Sensor 1 (dissolved oxygen) Sensor 2 (pH/ORP) (note) Fault 5.4
High or low High High (note) NA 5.4
Deadband 0 0 NA 5.4
Setpoint 0 ppm (high); 20 ppm (low) 0 ppm (high); 20 ppm (low) NA 5.4
(Dissolved oxygen - ppm)
Setpoint 0 ppb (high); 990 ppb (low) 0 ppb (high); 990 ppb (low) NA 5.4
(Dissolved oxygen - ppb)
Setpoint 0 ppm (high); 200% (low) 0 ppm (high); 200% (low) NA 5.4
(Dissolved oxygen - % sat.)
Setpoint (pH) 14 (high); 0 (low) 14 (high); 0 (low) NA 5.4
Setpoint (ORP/Redox) 1400 mV (high); -1400 (low) 1400 mV (high); -1400 (low) NA 5.4
Note: For single sensor input, alarm 2 is assigned to sensor 1, and is configured as a low alarm.