Instruction manual
4.5.7 ADJUST mA Menu
The analog output signals are factory-adjusted (upper limit: 20mA). However, if the upper limit of one of the
outputs, drifts, the span value can be adjusted with the mA ADJUST menu. Connect an ampere meter in series
to the analog output terminals. Adjust the value until the ampere meter displays 20mA.
ADJUST mA
Select
OUTPUT 1
OUTPUT 2
Enter
OUTPUT 1/2
VALUE: 0000
- +
Select
SERVICE
IMPEDANCES
AVERAGE
DISPLAY
CODE
SOFT ISSUE
DEFAULT VAL.
ADJUST mA
Enter
Model 14 pH/Redox Transmitter