Specifications
e-mail:
support@broadcasttools.com
voice:
360.854.9559
fax:
866.783.1742
19
Site Sentinel
®
16 Installation and Operation Manual
SETUP
Temperature
Temperature Label: Used to identify the temperature probe.
Alarms High: This option enables the “High Trip Point” email alarm.
Alarms Low: This option enables the “Low Trip Point” email alarm.
Alarms Exit: This option enables the emailing after exiting an alarm.
Email Addresses: This allows you to configure up to 8 emails to be sent when-
ever the input goes above or below a set-point.
Log Device: This enables the email logging of this device.
Delay time: The delay is in (0 to 999) seconds. This option specifies the
wait time from when a value is out of range before an alarm
is activated.
Low Temp: This option specifies the LOW alarm set point.
High Temp: This option specifies the HIGH alarm set point.
Temp Type: This option specifies what scale will be displayed, Fahrenheit
or Centigrade.
Hysteresis: Hysteresis (deadband), This option specifies the hysteresis
used when evaluating alarm conditions.
NOTE: Hysteresis prevents alarms from toggling excessively when temperature
is at the set point. This is due to normal fluctuation. For example, if the hystere-
sis is set to 1 degree, and a high alarm is to occur at 91 degrees, the hysteresis
ensures that once the high alarm is triggered, it won't go off until the temperature
returns to below 91 degrees (90 - 1). This reduces problems with small fluctuations
in temperature readings triggering multiple alarms. The hysteresis is also used
during low alarm conditions. If, for example, a low alarm is to occur at 43 degrees
and the hysteresis is set to 1 degree, then once the low alarm is triggered, it won't
go off until the temperature returns to higher than 44 degrees (43 + 1).
NOTE: After any item has been changed, you MUST press the “Save Settings”
button for your changes to be saved.