User's Manual

MJ192 GB Issue 02
4
MAIN MENU – QUICK REFERENCE
TEMP
OK Å Æ
RH%
OK Å Æ
ALARM HI
OK Å Æ
ALARM LO
OK Å Æ
C/F
OK Å Æ
SAVE
OK Å Æ
CANCEL
OK Å Æ
37.5C
OK - +
RH 20%
EX ONLY
HI 2.0C
OK - +
LOW 3.0C
OK - +
DISP C
OK C F
INCUBATION TEMPERATURE.
RANGE 20.0 – 40.0°C (68.0 – 104.0°F).
DEFAULT 37.5°C (99.5°F).
RELATIVE HUMIDITY.
RANGE 20% – 80%.
DEFAULT 20%.
ONLY FOR EX VERSION WITH
ADVANCE HUMIDITY PUMP – SEE
SECTION 8.
HIGH TEMPERATURE ALARM.
RANGE 1.0 – 5.0°C (1.8 – 9.0°F) ABOVE
SET INCUBATION TEMPERATURE.
DEFAULT 2.0°C (3.6°F).
LOW TEMPERATURE ALARM.
RANGE 1.0 – 5.0°C (1.8 – 9.0°F) BELOW
SET INCUBATION TEMPERATURE.
DEFAULT 3.0°C (5.4°F).
CELSIUS / FAHRENHEIT DISPLAY.
SWITCHES ALL TEMPERATURE
FIGURES BETWEEN °C AND °F.
DEFAULT °C.
SAVE.
ALL CHANGES ARE SAVED. RETURN
TO NORMAL OPERATION SCREEN.
CANCEL.
ALL CHANGES ARE IGNORED.
RETURN TO NORMAL OPERATION
SCREEN.
TURN 0/1
OK Å Æ
TURN INT
OK Å Æ
TURN 1
OK - +
MINS 90
OK - +
AUTOMATIC TURNING STATUS.
SWITCHES THE AUTOMATIC
TURNING SYSTEM ON (1) OR OFF (0).
DEFAULT OFF (0).
TURNING INTERVAL.
SETS THE DELAY BETWEEN TURNS.
RANGE 30 – 180 MINUTES.
DEFAULT 45 MINUTES.