16.3.4 PurposeCheck Pressure
MaterialNA
Manual toolWrench
NoticeNA
PM StepZero value confirmation Remove the XCDA, N2, and Exhaust
pipes, and observe whether the meter head value is zeroIf the value is not
zero, please reset to zero
XCDA, N2, Exhaust pipeline position, please
use a wrench to remove
At the same time the pressure will be
reset to zero by pressing the left and
right buttons for 3 seconds