Guidance and manufacturer’s declaration – electromagnetic emissions
The PRT2000 equipment is intended for use in the electromagnetic environment speci ed
below.
The customer or the user of the PRT2000 should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfer-
ence in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions
IEC 61000-3-2
Not Applicable
The ME equipment is
solely battery powered.
Voltage  uctuations /
icker emissions
Not Applicable
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
d = [
3.5
] P
V
1
80 MHz to 800 MHz
d = [
3.5
] P
E
1
800 MHz to 2.5 GHz
d = [
7
] P
E
1
0.01 / 0.12 0.23
0.1 / 0.38 0.73
1 / 1.2 2.3
10 / 3.8 7.3
100 / 12 23
Guidance and manufacturer’s declaration – electromagnetic immunity
The PRT2000 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the PRT2000 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environ-
ment – guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete, or ceramic tile.
If  oors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside
the shielded location
from  xed RF transmitters,
as determined by an
electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Separation distance
calculation provided above.
If a known transmitter is
present the speci c distance
can be calculated using the
equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not Applicable
(no electrical
cabling)
Electrical fast
transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment
is solely battery powered.
Surge
IEC 61000-4-5
±1kV
di erential
±2kV common
Not Applicable
Power frequency
magnetic  eld
IEC 61000-4-8
3 A/m Complies
Power frequency
magnetic  elds should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
>95% dip 0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Not Applicable
The ME equipment
is solely battery
powered.