Specifications
22
Specifications
Model: BP 1750 / BP 1700
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0–320 mmHg (cuff pressure)
40–199 / min (pulse rate)
Blood pressure
measurement: 20 mmHg (minimum diastolic value)
300 mmHg (maximum systolic value)
Laboratory accuracy: ± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy: According to AAMI-SP10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Inflation: Automatic
Deflation: Step valve
Power supply: 2 batteries, type LR 03 (AAA)
Operating temperature: + 10 °C to + 40 °C
Storage temperature: – 20 °C to + 60 °C
Humidity: Up to 85% RH
Cuff: Suitable for wrist circumference of 13 to 21 cm
Clinical validation: According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to different results for devices
which are calibrated against auscultatory reference.)
If device is not used within specified temperature and humidity ranges the technical
accuracy can not be ensured.
Type BF equipment
Subject to change without notice.
This device conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety