Specifications
DL900 Holter Monitor
7
3-Channel (5 Lead) Electrode Placement (1
st
option)
5 color-coded leadwires are used to create a 3-Channel ECG
recording. This is a typical electrode placement. Refer to your
Analysis System software and the physician for a recommended
position.
Figure 3 – 5-Lead Electrode Placement
# Channel Color Location
1 3- White
Next to the right Manubrium border on
the Clavicle
2 1-, 2- Red Centered on the Manubrium
3 2+, 3+ Black Lower left rib margin over bone.
4 1+ Brown Left Anterior Axillary line on the 6
th
rib
5
Green Lower right rib margin over bone.
DL900 Holter Monitor
8
3 Channel (5-Lead) Electrode Placement (2
nd
option)
5 color-coded leadwires are used to create a 3-Channel ECG
recording. This is a typical electrode placement. Refer to your
Analysis System software and the physician for a recommended
position.
Figure 4 – 5-Lead Electrode Placement
# Channel Color Placement
1 3- White
Right side below the V1 position, at
the bottom of the rib cage
2 1-, 2- Red
Center on the Manubrium, the top of
the sternum
3 2+, 3+ Black Left side at the V5 position, on a rib
4 1+ Brown Left side at the V3 position, on a rib
5
Green Right side opposite V5 position.