Specifications

DL700 Holter Monitor
7
Product Operation
DL700 Holter Monitor
8
3. Recorder Preparation
3.1 Battery Installation
To ensure a full 24 hour recording it is essential to use new batteries before
each patient recording. The recommended batteries for the DL700 Holter
recorder are two AA alkaline batteries (IEC-LR6). When inserting the
batteries, follow the Start Recording procedure in the Operation section
while making sure to follow the polarity indication shown on the label in the
battery compartment.
- Note -
There is a built-in polarity guard feature to protect against damage to the
recorder but if the batteries are installed improperly, the unit WILL NOT
RECORD and the LED WILL NOT BLINK.
3.2 Setting the Clock
Time and Date on the digital clock can easily be set by the following
procedure using the two small access holes labeled "set" and "select". The
holes are located next to the Patient Event Button shown in Figure 1.
a) Insert and press a pointed object into the top access hole, labeled "set",
to start the setting process.
b) Press the pointed object into the bottom access hole, labeled "select".
The digital display alternately shows the time and date.
c) Press the bottom access hole a second time to select MONTH.
- Note -
Pressing the bottom access hole repeatedly, cycles the digital clock
through each selection - Month, Date, Hour, Minutes.
d) To set the current month, press the top access hole repeatedly to obtain
the desired month.
e) Press the bottom access hole again to select DATE.