Specifications

DL700 Holter Monitor
19
Problem:
Noisy ECG Recording
Solution:
- Ensure that all the electrodes are firmly and securely attached to
the patient. (Refer to “Electrode Attachment” section).
- Replace the entire lead set. (These leads are frequently broken
inside the lead insulation due to someone pulling on the leadwire
instead of at the base of the lead to remove it from the patient input
connector).
DL700 Holter Monitor
20
8. Service Items
Part Number
1 HeartStor card, 16 Mbytes 350-0147-01
2 HeartStor card, 32 Mbytes 350-0147-02
3 HeartStor card, 64 Mbytes 350-0147-03
4 Digital clock module 800-0364-01
5 Lead set, patient, 7 lead 200-1380-001
6 Lead set, patient, 5 lead 200-1380-002
7 ECG Test cable box 200-1588-001
8 Pouch with strap and belt 200-1508-001
9 Filler plugs 200-1586-001
10 Shoulder strap, pouch 100-1693-001
11 Belt strap, pouch 100-1694-001