Specifications
DL700 Holter Monitor
17
Figure 3
Replacing Clock Battery
d) Remove the used battery and dispose of it properly.
e) Position the new battery in place with the + side facing the direction as
shown in Figure 3.
f) Reposition the metal holder and reinstall the first screw. Tighten the
two screws. Re-insert the clock module into the recorder.
g) Reset the time and date. Refer to the Operations section of this manual
for information on setting the clock.
DL700 Holter Monitor
18
7. Troubleshooting Guide
If there is a problem with the recorder, first review the “Typical Problem
Descriptions and Recommended Solutions” listed below.
If assistance is needed, call the following number for customer support
and/or arrangements for service of the recorder:
Braemar, Inc.
1285 Corporate Center Drive, Suite 150
Eagan, MN 55121 USA
Phone: 800.328.2719
Fax: 651.286.8630
E-mail: service@braemarinc.com
Web: http://www.braemarinc.com
7.1 Problem Description and Recommended Solution
Problem:
Green LED is not blinking.
Solution:
- Ensure that a Flash Memory card of at least 10 MB capacity has
been properly inserted in the recorder.
- Install a new set of AA alkaline batteries which are known to be
fully charged. New batteries MUST be used for each Holter
recording.
- Verify that the batteries are oriented according to the label in the
bottom of the battery compartment.
- Reinitialize by lifting one of the AA batteries for several seconds
and then re-inserting.
Problem:
Digital clock is blank
Solution:
Replace clock battery by referring to the Maintenance section in this
manual.
- Note -
The digital clock is independent from the operation of the DL700 and
does not affect the recording.