Specifications

DL700 Holter Monitor
15
Service And Maintenance
DL700 Holter Monitor
16
6. Maintenance
The DL700 Holter Recorder requires virtually no maintenance. Ensure that
the battery contacts are clean and that the inside of the recorder is free of
dust and dirt. The case and the pouch with belt and strap can be cleaned
with a lint free towel moistened with mild soap and water. Do not use any
strong solvents. Clean as often as desired.
6.1 Replacing the Clock Battery
The digital clock uses one 1.5V silver oxide battery (Duracell 392B or its
equivalent). The battery has approximately a two year life. The battery is
accessed under the top cover.
a) Open the top cover. Use a small jeweler's screwdriver inserted into the
exposed center screw on the back of the clock module to gently lift the
clock out of the endcap.
Figure 2
Removing Clock
b) Using the jeweler's screwdriver, remove the top small screw in the
metal holder that keeps the battery in place.
c) Loosen the other screw holding the metal holder. Swing the metal
holder to the side.