Installation / Operation Instruction Manual

47
Troubleshooting
RTG-K-199/160N/X2 – 6 720 813 642 (2022/06)
In case of failure in the auto calibration process the error sign
appears.
Verify if:
Any error codes appear on the display. See table 25.
Gas type and pressure are correct.
Wind conditions are stable (e.g. wind gusts may be a cause
of failure).
Hot water flow is sufficient and stable (e.g. water from a
well is a potential source of water instability).
In case of positive results for all these actions:
Exit UC menu and repeat the process.
Perform manual adjustment according to chapter 8.11.2
or contact your original installer or a local certified gas
technician for further assistance.
8.11.2 Manual adjust Gas/Air flow
Gas/Air flow adjustment is also possible to perform manually
(only recommended if automatic adjustment can not conclude
successfully).
First step - Adjust the maximum power (P1)
Press the button to turn ON.
Press and hold simultaneously buttons , and
for 3 seconds, until display reads P4.
Fig. 46
Press the buttons or to select P1 (Maximum
Power) menu.
Open hot water faucets when the display shows Tap
Symbol image (1 tub and 2 sinks should be sufficient;
when there is enough water flow the image disappears).
Press to enter P1 adjustment.
Display shows E.
Press or until L1 appears.
Press the button to enter L1.
Display shows L1 value.
Verify if power bar level is at middle point, see fig. 47.
Fig. 47 Power bar middle point
Press the buttons or to adjust parameter value if
power bar level is out of range.
1)
Fig. 48 Gas adjust - L1 parameter
Press the button for 3 seconds and appliance assumes
value.
Press the button to save the value.
Display will be back to P1 menu.
Press the button to enter L1 and verify if power bar
level is at middle point, see fig. 48.
Repeat the process until L1 gas adjust is correct.
Press button to finish the L1 gas adjust.
Press the button or until the display shows "A1".
Press the button to enter A1.
Display shows A1 value.
Verify if power by level is at middle point, see fig. 47.
1) Increasing L1 value will cause an increase in power bar level
and vice-versa.