Instruction manual
24 
WEIGHT UNIT kg
"* 
CAPACITY 15 
" (Input capacity)   
Input capacity of scale 
"* 
INC.0.002/0.005 
"Select interval with up arrow key [ ]↑     Select interval of the scale 
"*   
CAPTURE 0 ———   Read zero 
"*   
ADD WEIGHT 0.000    
"Input the standard weight 
value you want to put on scale 
  Input the standard weight value you want to 
calibrate, at least 2/3 of capacity 
"put on the standard weight on 
platter 
"*  
DONE 
SHUT DOWN
  Power off scale and power on again 
NOTE:
IN THE PROCESS IF GIVE THE MESSAGE “ INIT. CELL...”, WHICH INDICATE THE FAULTCALIBRATION. 