User’s manual V300+ / V600+ / V1000+ Version N°2.0 – E TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland Page 1 Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland Page 2 Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.
Contents 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. Description Page Description of the instrument Display functions Distintive characteristics Specifications Delivery Unpacking and Installation Getting started Replacing the battery pack Ajusting the measuring pressure Ajustement of the floating probe suspension Main functions Secondary functions How to start up 13.1 Setting the probe constant 13.2 Measuring heights 13.3 Measuring diameters and centerlines 13.
35 34 33 32 44 31 36 37 38 39 45 40 41 42 43 6 10 9 8 7 17 18 19 30 6 29 12 11 3 2 1 5 4 28 20 27 21 26 25 24 22 23 14 13 15 15 TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland Page 4 Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.
1. Description of the instrument 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. Handwheel allowing the displacement of the measuring probe. Locking device to activate the fine ajustment. Fine adjustment screw. Switch to acivate the air cushion. (Version C) Handle allowing the displacement of the instrument. Measuring probe. Screw to adjust the measuring pressure (0.7N-1.6N, ajustable) (wrench /2mm) Lever for fast displacement.
3. Distintive characteristics The V300,V600 and V1000 are height measuring instruments, self-contained, controlled by a microprocessor. o Easy to operate. o Self-contained operation for a period of: without air cushion : max.100h. : with air cushion : max. 50h. o Rechargeable battery. o Big display. o Acceptance of measured values by an acoustic signal. 4. Specifications Designation Measuring range Application range Resolution Overal measuring accuracy Repeatability Max.
Options : see optional accessories (page 22-26) TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland Page 7 Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.
6. Unpacking and installation The V300, V600 and V1000 are supplied in a shock prof box. The instrument itseff is surrounded with a protection cover. - To unpack the instrument lay the box in a horizontal position, Brand name visible on top, open it and take the instrument out carefully. - Take off the protection cover. Insert the ruby ball (V-1). Unlock completely the transport safety screw (9), Screw, chromium-plated. Clean the instrument, especially the three point air bearings on the base.
7. Getting started The instrument stands on a granite surface table (or e.i. cast Iron table) - Switch ON the instrument by pressing the key (ON). - The display will show rEF I MM - The reference of the column must be taken : - Move the probe (A) slowly and cross A the two arrows to take the reference. This operation will memorized this reference point improve the accuracy of the instrument. - The display will begin to count. If during this operation the display doesn’t count. restart the operation again.
8. REPLACING THE BATTERY PACK As sson as the autonomy of the instrument becomes unsufficient, the battery pack should be changed.(Lifespan about 4 to 6 years) 1. 2. 3. 4. 5. 6. Purchase a battery pack from your TRIMOS distributor. Switch off the instrument. Place the instrument in an horizontal position onto a table. (steady) Unscrew the base plate. (5 screws) Disconnect the battery pack (A) Connect the new battery pack and refit the base plate. Take care not to damage Any connecting cables. 7.
9. Adjusting the measuring pressure DO NOT ADJUST UNLESS IT IS REQUIRED ! Tools to be used - 1 Hex screwdriver wrench ( 2 mm) - 1 Dynamomètre (force gage) 0 – 300 grs T016-INB.2 - The adjustment screw (7) is situated at the left side of the column, inside on the measuring carriage. - The measuring pressure is factory set at approx. 1 N. - By turning this adjustment screw anticlockwise, the measuring pressure will be decreased.(2 turns = ~0.
11. Main functions Switches the instrument ON or OFF. Sending data RS232, only when measuring. ((See page 18) for other settings) Set the display at zero or memorizes preset values. Normal key mode, green light on, measuring heights of normal and Reversed surfaces. (The probe constant must have been previously entered to measure reversed surfaces. Æ , key mode, orange light on, measuring diameter (takes account of probe constant). Measures internal or ecternal diameters in any sequences.
12. Secondary functions Memorizing a preselected value : Press Preset key (long pressure on the key ~2 sec.) Introduce a preselected value from the numerical Keyboard. Confirm the preselected value by pressing again The PRESET key. The preselected value is memorized. (A second preselected value can be memorized on REF II mode.) 00000000 MM PRESE T 0123.000 MM PRESE T 0123.000 MM Initialization functions : + + To lock or unlock the conversion : switch the instrument On while pressing the mm/in key.
13. How to start up ? 13.1 Setting the probe constant. To measure reversed surfaces, or diameters. the instrument takes account of the probe constant, i.e. the diameter of the ball and its deflection. The green light comes on when the probe comes into contact with the part being measured. P1 P2 1. Switch ON the instrument, then press key Move the probe across the two arrows to initialize the column reference (see page 8) REF I MM 0.123 MM 2. Memorizing the constant : press 0.123 key MM 3.
13.3 Measuring diameters and centerlines. The probe constant value should already have been memorized : (see page 13) 1. Normal mode, if necessary press key, lights OFF 0.000 MM 2. Touch the reference plane (P1) green light ON 3. Press PRESET key quickly (>1 sec.), to set the display at zero. 21.000 M M 0.000 Realise the probe. MM 4. Change to Diameter/Centerline mode by pressing key, orange light blink. 0.000 ∅ P3 31.003 MM 5.
13.4 Measuring with min and max mode. 1. Normal mode, if necessary press Key (lights OFF) 0.000 MM 2. Touch the reference plane (P1) (green light ON) 3. Press PRESET key quickly (>1 sec.) to set the display at zero. key 24.456 M M 0.000 MM 4. Place the probe inside the groove to be measured and touch a point (P2) which is well away from the reversal point. 5. Press MIN/MAX key to MIN MIN 125.231 6. Move the part or the instrument sideways to the probe. The reversal point is memorized.
13.5 Measuring with Delta mode P2 P1 1. Normal mode, if necessary Key (lights OFF) 0.000 MM 2. Touch the reference plane (P1) 102.678 MM 3. Delta mode, press MIN/MAX key to 0.000 (Press 3 times the key) (MIN,MAX,DELTA) MM 4. Move the part or the instrument sideways to the point (P2). Delta is memorized. (Delta = maximum-minimum) 0.022 MM 13.6 Measuring of squareness. 1. Use a lever dial indicator and clamp it into the holder (A). 2.
13.7 Measuring with two references 1. Normal mode, if necessary press Key (lights OFF) 0.000 MM 2. Touch the reference plane (P1(REF I)) -12.456 MM 3. Press the PRESET key To set the REF I to zero. P2 0.000 REF II MM 299.985 MM 5. Press REFII -184.912 4. Touch the reference plane (P2(REF II)) P3 130.098 Key to activate the REF II MM 115.073 6. Press the PRESET key To set the REF II to zero REFII 0.000 P1 REF I MM 7.
14. Initialization of printing mode + Instrument OFF Hold down the PRINT key when switching ON Pri 1 MM by pressing PRINT key again. Pri 2 MM by pressing PRINT key again. Pri 3 MM When your choice of printing mode is done, press any numerical key. nr oFF MM nr on by pressing the PRINT key. MM To exit of the initialization printing mode press any numerical key. 123.
15. Printer Introduction The thermal printer V-30 is compact and light weight, equiped with an RS232C serial interface via a 9 –way D-type connector (female). It is powered from internal Ni-Cd batteries. Material required : Printer with AC adaptor (V30/…), Bracket (V32), connecting cable (V-31), paper rolls (V-30.7). Assembling printer on V300,V600,V1000. 1.
16. AC adaptor, external contact If the indicator (36) showing complete discharge is displayed, connect the AC adaptor. The instrument remain operational. Complete recharding takes 15 hours. A foot pedal may be connected to the external contact (16) for data transfer. Another functions may be assigned to it.. (see page 21) 17. RS232C connector The RS 232 C port (13) is OptoRS compatible (see OptoRS convention for more informations).
18. Remote control of the instrument The instrument functions are controlled externally by an ASCII code corresponding to the 3 first letters of the function. Spaces are removed. The message may be in upper or lower case. The question mark ? implies an instrument response. ? or PRI (PRInt) CLE (CLEar) requests displayed value. reinitialises the maximum / minimum / delta memories. Code for external functions.
3’’ Setting/reference gauge, H = 75mm V75 Setting/reference gauge, H = 3 in. V75E 75 19. Optional accessories Tungsten-carbide ball probe, Ø 4mm V-1.4/L120 22 72 8-h6 V-1 4 Ruby-ball probe (standard), Ø 4mm 4 8-h6 12’’ 300 Setting/reference gauge, H = 300 mmV-300M Setting/reference gauge, H = 12 in.
8 19. Optional accessories V-7/H100 Fast locking device for measuring probe. V-20 8-h6 100 Extension probe holder wooden accessories support V-60 (without accessory) 20 TVA 4 Holder TVA5.1 and Probe holder TVA5.2 TVA 5M Holder TVA5.1 and Probe holder TVA5.2E TVA 5E M2.5 8-h6 4/48’’ 95 4/48’’ 8-h6 2.2 95 R 15 TVA 9.1 10 Barrel-shapped insert (M6-M48) TVA 9.2 Barrel-shapped insert (M12-M150) TVA 9.3 8-h6 55 4.5 R 65 22 9.7 60 8-h6 Barrel-shapped insert (M3-M16) 8-h6 M2.
4 19. Optional accessories Ball insert TV 7.2 Pin shaped insert TV 7.3 25 4 TV 7.1 3 Knife-edge insert 2 4 24 Disc shaped insert 4 24 5 8 TV 7.4 25 Complete set (set N°1) including : Ruby-ball probe Ø 2mm, L=87mm Holder (V-50.12.1) and probe Ø 1mm (V-50.12.2) Cylindrical probe Ø 3mm, L=89mm Knife-edge insert L= 89mm Locking key for probe Hex screw wrench 1,5mm Hex screw wrench 2,5mm Wooden case V-51 Complete set (set N°2) including : Ruby-ball probe Ø 2mm, L=87mm Holder (V-50.12.
19. Optional accessories Printer including : Printer (V30.0) and AC adaptor (without bracket, without cable) V-30/110US V-30/220 V-30/240/GB Bracket including 3 screws V-32 RS 232C cable V -> printer V-31 RS 232C cable V -> PC CABL.RS.1/1-9P Thermal Paper rools (5 pcs) V-30.7 AC adaptor for printer V30/… AC adaptor 110V/US AC adaptor 220V/Europe AC adaptor 240V/GB V-30.1/110US V-30.1/220 V-30.
20. Maintenance Remember that it is a measuring instrument ? Take care of it. 21. Transport For re-packing, use the original packing or adequate packing. Protect the instrument by a slipcover. Do not put accessories close to the instrument. 22. Complaints / Repairs In case of breakdown, contact the distributor of TRIMOS immediately. Repairs can only be considered if the instrument is sent to us in a clean condition and Packed according to our instructions.
23. Distributors worldwide South Africa Austria TRIMOS SYLVAC SA PTY LTD P.O.Box 95672 Waterkloof 0145 PRETORIA ' 0027 12 6530747 Fax 0027 12 6530749 UPM Handelsges.m.b.h. Rupertiweg 5084 GROSSGMAIN ' 0043 6247 8229 Fax 0043 6247 8696 E-mail upm@salzburg.co.net Germany PRECIMA GmbH August-Schanz-Str. 28 60433 FRANKFURT/M ' 0049 69 954326-0 Fax 0049 69 954326-99 E-mail info@precima.de Web www.precima.de Argentina ARO S.A. Av.
23. Distributors worldwide China - Hong Kong DANTSIN TECHNOLOGIES LTD 7B Greenery Court Discovery Bay - HONG KONG ' 00852 29878852 Fax 00852 29878865 E-mail dan@ibm.net China - Hong Kong WYLDAR MACHINE TOOL LTD Fabrco Ind. Bldg., 1/F 78-84, Kwai Cheong Rd KWAI CHUNG N.T. ' 00852 2 4899881 Fax 00852 2 4804721 South Korea TSM CORPORATION # 602 Heung-II Building, 1073, Hogae-Dong, Dong-an Ku ANYANG CITY / KYUNGKI-DO ' 0082 343 3886724 Fax 0082 343 3886727 E-mail tsmc@unitel.co.
23. Distributors worldwide Indonesia CV. UNI TECHNIC Jln. Naripan n° 85 BANDUNG 40111 ' 0062 22 439800 Fax 0062 22 437600 E-mail wijaya@ibm.net Japan TRIMOS SYLVAC METROLOGY (JAPAN) LTD 16-22 Karahoricho Tennoji-ku OSAKA ' 0081 6 67614281 Fax 0081 6 67612150 E-mail tsmosaka@apricot.ocn.ne.jp Jordan Indonesia PT.UNI METRIKA UTAMA Jl.Agung Niaga 3 Block GIII N°48 Jakarta Utara 14350 ' 0062 21 645 2201 Fax 0062 21 640 4680 E-mail umu1@dnet.net.
23. Distributors worldwide Poland TRIMOS SYLVAC S.C. ul. Pustelnicka 48 m.10 04-138 WARSAW ' 0048 22 6123446 ou natel 602650825 Fax 0048 22 6123446 Portugal EUROTECNOLOGIA Maquinas e Equipamentos Industriais, Lda Rua Delfim Ferreira, 739 4100 PORTO ' 00351 22 6173004 Fax 00351 22 6173801 U.K. and Ireland BOWERS METROLOGY UK LTD Unit 7 Crown Business Centre Horton Road, West Drayton MIDDLESEX UB7 8HZ ' 0044 1895 430001 Fax 0044 1895 430002 E-mail sales@bowers.co.uk Web www.bowers.co.
23. Distributors worldwide Turkey TEKNO 2000 A.S. Incirli Cad. No: 37/3 34740 BAKIRKÖY/Istanbul ' 0090 212 5424202 / 5438322 Fax 0090 212 5424363 E-mail info@tekno2000.com.tr Web www.tekno2000.com.tr Russia FRIBOSA AG Werkzeug- und Maschinenbau Freiburgerstrasse 71 CH - 3280 MURTEN ' 0041 26 678 78 78 Fax 0041 26 678 78 70 Vietnam DATH SINH CO. LTD. 97 Hoang Si Khai St. 8th Dist, HCMC ' 0084 8 8555232 Fax 0084 8 8555232 E-mail lindalienbach@yahoo.com Yougoslavia TEHNOALAT EXPORT-IMPORT sa p.o.