Instruction Manual
PC-2200
BOWERS METROLOGY page 5 /32 version 2.06 -E
1.2 TECHNICAL SPECIFICATIONS
General :
Storage temperature : -40°C to 60°C /- 40°F to 140°F
Operational temperature : 0°C to 50°C / 32°F to 122°F
Power supply : 110/230V (AC) 50/60Hz
Max. consumption : 40VA
Fuse : 2A delayed
MTBF (Mean Time between Failure) : >100'000 hours
Average life of display : > 50'000 hours
Performance :
Time for stabilization after switching ON : 5 minutes
Measurement thermal drift : 100PPM/°C
Measuring thermal drift between channels : 30PPM/°C
PC-2200 equiped with HBT-LVDT-CAP probes : ± 0.5% ± Resolution
Measuring accuracy at 20°C / 68°F
Resolution = 0.1 µm / .00001’’
PC-2200 equiped with Air System
in the nominal conditions
Working range : ± 50µm / .002’’
Interfaces :
Analog OUTPUT
Range : ± 5.0V F.S.
Full scale accuracy: : 2% of full scale
Serial OUTPUT (RS 232)
Type : RS 232 C
Transmission speed (Baud rate) : 4800, 9600, 19200
Word length : 7 or 8
Parity control : NONE, EVEN, ODD
DC-OUT Photo couplers
Maximal Voltage : 50V
Maximal Tension : 60mA
Weight and Dimensions :
Dimensions (height * width * depth) : 450 * 57 * 215mm / 17.7’’*2.16’’*8.66’’
Weight : 4.12 kg